Clarinet (Clinical Middle Ear Analyzer)

Clarinet is a clinical middle ear analyzer, featured with all the tests needed for a deep and accurate investigation of middle ear disorders. Automatic and manual tympanometry, acoustic reflex threshold, decay and latency examinations, ETF for intact and perforated eardrums, high frequency and multi-component tympanometry are all available and full results can be observed on a wide color touch screen display.

The display

Clarinet incorporates a large high-resolution color graphic display, giving a clear and detailed picture of the exam in progress. And with touch screen input, interaction with the instrument is extremely simple.

The probe

Clarinet features an extremely light, small and sturdy metal probe, which can be very rapidly disassembled and cleaned. The probe is connected to the small Control Box, through which you can select the ear and start the exam.

Endless flow pump

Clarinet immediately impresses for the incredible speed of its pump: the sweep speed can reach 600 daPa / sec, without compromising on accuracy and precision of tympanogram plots. The pump features an endless airflow technology, improving the execution of the test when the seal is not optimal.

Go in-depth with investigations

High-frequency and multi-component tympanometry, evaluation of multiple stimuli averaged reflex latencies together with the up-to-date measure of reflex plot angles of incidence are just some of the Clarinet advanced features that will take your clinical investigation to new levels.

Noah & Maestro integration

Simply connect Clarinet to a USB port on your computer (no driver needed), and you can transfer exam results into your Maestro or Noah database. And with the live view feature, you can have a wide view of the examination in progress on the computer display.

 

PROBE TONE 226 HZ AT 85 dB SPL
678HZ, 800HZ, 1000HZ (PLUS)
COMPL. RANGE 0.2 TO 8.0 ML / 0.9 TO 15.0 MMHO
PRESSURE RANGE FROM -600 TO +400 daPa
PUMP CONTROL AUTOMATIC AND MANUAL
PRESS. CHANGE RATES 15, 50, 100, 200, 300, 400, 600 daPa/sec, AUTO (200-600)
PRESS. SWEEP BIDIRECTIONAL
TYMPANOMETRY MEASURES ADMITTANCE [Y]
SUSCEPTANCE [B] (PLUS)
CONDUCTANCE [G] (PLUS)
ACOUSTIC REFLEX TESTS MANUAL TEST
AUTOTHRESHOLD
FIXED INTENSITY
GROWING INTENSITY
REFLEX DECAY
REFLEX LATENCY
STIMULATION IPSI AND CONTRALATERAL
STIMULI 250 (CONTRA), 500, 1K, 2K, 4K, 6K, 8K (CONTRA) HZ, BBN, HPN, LPN
MAXIMUM INT 110 DB HL (IPSI); 120 DB HL (CONTRA)
STIM. DURATION FROM 0.5 TO 2.5 SEC (SELECTABLE)
REFLEX DECAY: 10 OR 20 SEC
AUTOMATIC TESTS QUICK A AND B (SEQ. OF TYMPAN. AND REFLEX)
ETF TEST INTACT OR PERFORATED EARDRUMS
MEMORY 3 TYMPAN. + 16 REFLEX + 4 DECAY + 4 LATENCY RECORD. PER EAR
DISPLAY TYPE COLOR GRAPHICAL WITH TOUCH SCREEN
DISPLAY SIZE, RESOLUTION 7”, 800 X 480
PRINTER BUILT-IN THERMAL PRINTER (OPTIONAL)
LANGUAGE MULTILINGUAL INTERFACE
PC INTERFACE USB (DRIVERLESS)
COMPATIBLE SOFTWARE INVENTIS MAESTRO OR NOAH WITH DEDICATED MODULE
DIMENSIONS 32 X 32 X 15 CM - 12.6 X 12.6 X 5.9 IN
WEIGHT 2.2 KG / 4.9 LBS
STANDARDS TYMPAN.: EN 60645-5 / ANSI S3.39, TYPE 1
SAFETY: EN 60601-1
EMC: EN 60601-1-2
CLASSIFICATION CLASS IIa (MDD 93/42)